GE HEALTHCARE
D
IRECTION 5141177-100, REVISION 14 BRIGHTSPEED ELITE, EDGE, EXCEL: PRE-INSTALLATION
Page 134 Section 3.0 - Interconnect Runs, Wiring and Cables
Section 3.0
Interconnect Runs, Wiring and Cables
3.1 GEMS Supplied (Optional, Short Run)
RUN #
LENGTH,
ACTUAL
(USABLE)
UL CABLE INFORMATION
UL Style
Flam.
Rating
Voltage
Rating
Voltage
Actual
Temp.
Rating (C)
Dia.
mm (inch)
# of Cond
Size AWG
PULL SIZE
MM (INCHES)
ft m
PART # DESCRIPTION
050 28
(20)
8.5
(6.1)
2343529-2 HVDC, PDU to Gantry 2587 FT4 600 + & - 350VDC 90 19 (.751) 3 (2) 4
(1) 8
22 (0.87) Dia
051 28
(20)
8.5
(6.1)
2343530-2 HVAC, PDU to Gantry 2587 FT4 600 440Y/254 90 15.3 (.604) 4 14 11.2 (0.44) Dia
052 28
(20)
8.5
(6.1)
2343528-2 LVAC, PDU to Gantry 2587 FT4 600 208Y/120 90 13.8 (.542) 5 8 56.4 (2.22) Dia
053 65
(60)
19.8
(18.3)
2343531-2 LVAC, PDU to Console 2587 FT4 600 120VAC 90 12.3 (.483) 3 10 56.4 (2.22) Dia
054 n/a LVAC, Gantry to Table 1015 600 120VAC 3 14
055 28
(20)
8.5
(5.97)
2371450-2 Ground, PDU to Raceway 1284 VW-1
(FT-1)
600 0 105 15.5 (.608) 1 1/0 15.8 (0.62) Dia
056 68
(57)
20.8
(17.4)
2371450-4 Ground, Raceway to Console 1283 VW-1
(FT-1)
600 0 105 11.9 (.467) 1 2 12.2 (0.48) Dia
100 32.5
(20)
9.9
(6.1)
5120646-2 Signal, Gantry MSUB to PDU FT-4 300 <30VDC 80 11.2 (.440) 25 22 17 x 58 (0.68 x 2.30)
19 x 51 (0.75 x 2.01)
101 71
(60)
21.7
(18.3)
5120645-2 Signal, Gantry MSUB to OC FT-4 300 <30VDC 80 11.2 (.440) 25 22 17 x 58 (0.68 x 2.30)
19 x 51 (0.75 x 2.01)
102 71
(63)
21.7
(19.3)
2373436-3 Signal (LAN), Gantry to OC 1900 <30VDC 5.9 (.234) 8 24 15 (0.59) Dia
103 68
(60)
20.7
(18.3)
2117848-7 Fiber Optic, Gantry to OC N/A N/A 1 N/A 10 (0.39) Dia
104 n/a Signal, Gantry to Table FT-4 300 80 25 22
Table 9-2 GEMS Supplied Cables (Optional, Short Run) - UL Information