GE HEALTHCARE
D
IRECTION 5141177-100, REVISION 14 BRIGHTSPEED ELITE, EDGE, EXCEL: PRE-INSTALLATION
Chapter 5 - Environmental Conditions Page 99
5 – Environmental
Conditions
Section 2.0
Cooling Requirement
Use Table 5-1 to assist in cooling requirements planning. Over half the cooling used by your
scanner is required for gantry operation. Locate a wall air-conditioning vent at floor level beside and
behind gantry to meet both gantry cooling needs and provide patient comfort. Do not locate any
cooling vents directly above the gantry. Air returns above the gantry are recommended.
Please refer to Figure 4-7, and Figure 4-12 in Chapter 4, Section 6.0, for component air flow
requirements.
Section 3.0
Altitude
System operating altitude is from -492 ft. to 9842.5 ft. (-150 meters to 3000 meters).
SYSTEM COMPONENT BTU/HR WATT
1. Gantry recommended (See NOTE 1) 18,766 5,500
2. HPower Table / GT1700 Table 682/1023 200/300
3. Power distribution unit 5,118 1,500
Recommended Scan Room Subtotal (see notes): 24,566/24,907 7,200/7,300
4. Operator's console w/ 1 IG, 2 monitors & SCSI Tower 8,189 2,400
MOD Drive (Option) 682 200
IG Node, each (Option) 1,024 300
Remote Monitor (Option) 222 65
DASM (Option) 290 85
Recommended Control Room Subtotal (w/o option): 8,189 2,400
System Total (Recommended)
(See NOTE 1)
32,755/33,096 9,600/9,700
Option: Advantage Windows 256 75
Option: UPS 5,200 1,200
ROOM TOTAL (SEE NOTE 2)
NOTE 1:With 75 scan rotations per patient:
Recommended BTU/hr. provides for up to six patients per hour. It is also needed
during calibration of the system.
NOTE 2:Cooling requirements do not include cooling for room lighting, personnel or non-CT
equipment.
Table 5-1 Cooling Requirements (Worksheet)