EasyManuals Logo
Home>GE>Medical Equipment>Dash 5000

GE Dash 5000 Service Manual

GE Dash 5000
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
Configuration: Service menus
4-6 Dash 3000/4000/5000 2000966-542D
Service Mode options
Following is a list of options in the main code service menu.
Service Mode menu options
Menu option Description
REVIEW ERRORS For advanced troubleshooting by GE engineers. Error
log data can be transferred over the network to a central
station and then loaded onto a diskette for review. Refer
to “Error messages” on page 6-5 for details.
CALIBRATE For checkout or calibration of the noninvasive blood
pressure, ECG analog output, BP analog output, CO
2
service, and SAM service menu functions of the patient
monitor.
BATTERY SERVICE Provides battery data.
NOTE
If an INCOMPAT message displays in the Manufacturer Name
column and UNKNOWN displays in all other columns, the
installed battery is incompatible.
PATIENT-MONITOR TYPE For selecting the type of patient monitor desired, i.e
adult, neonatal or operating room. Refer to “Patient-
Monitor Type” on page 4-12 for detailed procedures.
WARNING
Changing the patient-monitor type will default the
admit function to Standard configuration. Different
alarms and parameters are activated for each selection.
NOTE
The keypad/remote control is DIDCA programmed for specific
patient monitor types. The error message, “WARNING:
REMOTE MISMATCHED WITH MONITORING MODE
displays if the patient monitor and keypad/remote control do not
match.

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals