Configuration: Service menus
2000966-542D Dash 3000/4000/5000 4-7
MENU SETUP Provides the following sub-menus: (Refer to “Service
Mode settings” on page 4-11 for detailed procedures.)
ADMIT MENU: STANDARD
Allows you to determine the function of the patient
monitor. The four variables include stationary or
ambulatory (telemetry) patient monitoring with a
monitor that always stays in one room (STANDARD)
or a monitor that moves from room to room (ROVER).
MONITOR DEFAULTS PASSWD
Allows you to set the patient monitor so that a
password is Required or Not Required for entry into
the Monitor Defaults menu section. If selected, the
password will be the same as the Service Mode menu
password.
MONITOR SETTINGS Provides the following sub-menus: (Refer to “Service
Mode settings” on page 4-11 for detailed procedures.)
SET UNIT NAME
Allows changes to the care unit name. After initial
setup, this name should not be changed or
communication to the central station will be corrupted.
Note that the care unit name must be registered
exactly the same in the central station and the patient
monitor.
SET BED NUMBER
Allows changes to the bed number. After initial setup,
this number should not be changed or communication
to the central station will be corrupted. Note that the
bed number must be registered exactly the same in
the central station and the patient monitor.
SET INTERNET ADDRESS
Allows changes to the internet (IP) address. An
incorrect internet address may also prevent the
patient monitor from viewing other monitors on the
network even though the unit names match. Whether
or not this can occur depends on the network topology
at the installed site.
WARNING
Duplication of an internet (IP) address on a network
causes lost data. If you change the factory assigned
internet address, you must first record all other internet
addresses used on your network to avoid duplication.
NOTE
An incorrect internet address may also prevent the patient
monitor from viewing other monitors on the network even
though the unit names match. Whether or not this can occur
depends on the network topology at the installed site.
Service Mode menu options (Continued)
Menu option Description

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals