Performing an Exam
2-52 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
Rev. 9
Open Trace
To trace the circumference of a portion of the anatomy and
calculate its length:
1. Press Measure key to enter into measure mode, if
necessary.
2. Adjust Caliper Change in the Primary Menu to Open Trace.
NOTE: The available values for Caliper Change menu can be set
in Utility -> Measure -> Advanced preset -> Measure Key
Sequence.
3. To position the caliper at the start point, move the Trackball.
4. To fix the trace start point, press Set. The caliper changes to
an active caliper.
5. To trace the measurement area, move the Trackball around
the anatomy. A dotted line shows the traced area.
6. To complete the measurement, press Set. The system
displays the circumference and the length in the Results
Window.
HINTS
Before you complete the trace measurement:
• To erase the line (bit by bit) back from its current point,
move the Trackball or adjust the Ellipse control
counterclockwise.
• To erase the dotted line but not the caliper, press Clear
once.
• To clear the caliper and the current data measured, press
Clear twice.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals