EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V1

GE LOGIQ V1 User Manual

GE LOGIQ V1
381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
Measurement and Analysis
LOGIQ V2/LOGIQ V1 – User Guide 2-57
Direction 5610736-100 Rev. 9
TAMAX and TAMEAN
Manual Trace
The value measured depends upon the Vol Flow Method preset.
The two selections available are: Peak (TAMAX) and Mean
(TAMEAN).
To do a manual trace of TAMAX or TAMEAN:
1. Press Measure key to enter into measure mode, if
necessary.
Adjust Caliper Change in the Primary Menu to S/D Trace.
In the Utility -> Measure - Advanced preset menu set Trace
preset to Manual.
2. To position the caliper at the trace start point, move the
Trackball.
3. To fix the start point, press Set.
4. To trace the maximum values of the desired portion of the
spectrum, move the Trackball.
NOTE: To edit the trace line, move the Trackball.
5. To complete the measurement, press Set. The system
displays the measurement values in the Results Window.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals