Getting Started
1-54 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
Rev. 9
Probes
Introduction
Only use approved probes.
Connecting the Probe
Probes can be connected at any time, regardless of whether the
console is powered on or off.
To connect a probe:
1. Place the probe's carrying case on a stable surface and
open the case.
2. Carefully remove the probe and unwrap the probe cord.
CAUTION
Inspect the probe before and after each use for damage or
degradation to the housing, strain relief, lens, seal, cable and
connector. DO NOT use a transducer which appears damaged
until functional and safe performance is verified. A thorough
inspection should be conducted during the cleaning process.
CAUTION
Remove any dust or foam rests from the probe pins.
CAUTION
Fault conditions can result in electric shock hazard. Do not
touch the surface of probe connectors which are exposed
when the probe is removed. Do not touch the patient when
connecting or disconnecting a probe.
CAUTION
DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals