Getting Started
1-22 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
Rev. 9
Peripheral/Accessory Connector Panel
LOGIQ V2/LOGIQ V1 peripherals and accessories can be
properly connected using the connector panel.
Each outer (case) ground line of peripheral/accessory
connectors are Earth Grounded.
Signal ground lines are Not Isolated.
For compatibility reasons, use only GE-approved probes,
peripherals, or accessories.
DO NOT connect any probes or accessories without approval
by GE.
The connection of equipment or transmission networks other
than as specified in these instructions can result in electric
shock hazard. Alternate connections will require verification of
compatibility and conformity to IEC/EN 60601-1 by the installer.
DO NOT touch the patient and any of the connectors on the
ultrasound unit simultaneously, including ultrasound probe
connectors.
DO NOT touch the conducting parts of the USB, Ethernet,
Video, Audio cables when connecting equipment to the unit.
When using peripheral device, observe all warnings and
cautions given in Peripheral manufacture’s manuals.