Performing an Exam
2-60 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
Rev. 9
M-Mode Measurements
Basic measurements that can be taken in the M-Mode portion of
the display are:
• Tissue Depth (Distance)
• Time Interval
• Time Interval and Velocity
NOTE: The following instructions assume that you do the following:
1. In the B-Mode part of the display, scan the anatomy you
want to measure.
2. Go to the M-Mode part of the display.
3. Press Freeze.
Tissue depth
Tissue depth measurement in M-Mode functions the same as
distance measurement in B-Mode. It measures the vertical
distance between calipers.
1. Press Measure key to enter into measure mode, if
necessary.
Adjust Caliper Change in the Primary Menu to MM
Distance.
An active caliper with a vertical and horizontal dotted line
displays.
2. To position the active caliper at the most anterior point you
want to measure, move the Trackball.
3. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
4. To position the second caliper at the most posterior point
you want to measure, move the Trackball.
5. To complete the measurement, press Set.
The system displays the vertical distance between the two
points in the Results Window.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals