XNX Universal Transmitter
Warnings/Faults
133
Notes
Note 2:
Spi Event Bits
Decimal Description
1
SPI1 Starting TX
2
SPI1 transmitting
4
falling clock edge, 0 = rising edge
8
SPI1 port open, 0 = closed
16
SPI1 no response
32
SPI1 ECC no response
64
SPI1 missing data
128
Not used
256
SPI3 Starting TX
512
SPI3 transmitting
1024
falling clock edge, 0 = rising edge
2048
SPI3 port open, 0 = closed
4096
Not used8192
16384
32768
SPI2 Starting TX
Note 3:
Subtypes Decimal Description
Fault 2
Event Bits
1 CRC error in safety critical RAM block
2 Error reloading safety critical RAM block from EEPROM
4 Error loading data from Personality board
8
Excel signal level has been below the low signal level
threshold for at least 24 hours
16 Excel beam blocked
32 Personality board error code > 0
64 Option board error code > 0
128 IR mA input > 1 mA and < 3.4 mA
256 IR mA input < 1.0 mA
512 IR forced 10 mA not within +/-1 mA
1024 gains from PGA don't match local copy
2048 error reading or writing EEPROM
4096 ECC reex failure
8192 RAM test failure
16384 Program memory CRC failure
32768 Op code test failure
Fault 3
Event Bits
1 Interrupt integrity test failure

Table of Contents

Other manuals for Honeywell XNX-UT Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XNX-UT Series and is the answer not in the manual?

Honeywell XNX-UT Series Specifications

General IconGeneral
BrandHoneywell
ModelXNX-UT Series
CategoryTransmitter
LanguageEnglish

Related product manuals