Front Panel Operation
3-122
Outputs used as logic inputs
To use the digital outputs as logic inputs to active TTL, Low-
power TTL, or CMOS inputs:
1. Connect the Model 2001 digital outputs to the logic in-
puts.
2. Connect the digital grounds.
3. Using the STATE menu, check the output state setting of
the Model 2001 output lines. The STATE value for each
output used should be ON.
4. Using the LOGIC-SENSE menu, check the logic-sense
setting of the Model 2001 output lines (TTL1 through
TTL4). Make sure the correct LOGIC-SENSE value is
Figure 3-48
Sample externally powered relay
To other three
digital outputs
+5V
10Ω
Digital Output #1
Flyback Diode
Digital I/O Receptacle
51
96
(Connector J1031)
Pin 4 - External Voltage Flyback
connection (+5V to +30V)
Pin 6 - Digital Output #1
Pin 5 - Digital Ground
Relay Coil
(+)
(-)
External Power
(+5V to +30V)
10kΩ
Pull Up Resistor
Relay Coil
(+)
(-)
External Power
(+5V to +30V)
Flyback Diode
Equivalent Circuit
Model 2001
Transistor Switch
selected for each output line. The LOGIC-SENSE value
varies according to the type of TTL, Low-power TTL, or
CMOS inputs used (ACTIVE-HIGH or ACTIVE-
LOW).
NOTE
If any LIMITS control is enabled
(LOLIM1 or 2, HILIM1 or 2—High, Low,
or Pass), the OUTPUT-STATE menu does
not check or change the output status.
Refer to paragraph 3.12.5 for information
about limits.