SCPI Command Subsystems
D-2
Table D-1
Calculate command summary
Command Description
:CALCulate[1]
:FORMat <name>
:FORMat?
:KMATh
:MMFactor <NRf>
:MMFactor?
:MBFactor <NRf>
:MBFactor?
:PERCent <NRf>
:PERCent?
:STATe <b>
:STATe?
:DATA?
:IMMediate
Subsystem to control CALC 1:
Select math format: MXB, PERCent, PDEViation, NONE.
Query math format.
Path to configure math calculations:
Specify “m” for mX+b (-1e21 to 1e21).
Query “m” factor.
Specify “b” for mX+b (-1e31 to 1e31).
Query “b” factor.
Set PERCENT value (-1e36 to 1e36).
Query PERCENT.
Enable (1 or ON) or disable (0 or OFF) calculation.
Query state of math function.
Read math result of CALC 1.
Recalculate input data.
:CALCulate2
:FORMat <name>
:FORMat?
:STATe <b>
:STATe?
:IMMediate
:DATA?
Subsystem to control CALC 2:
Select math format: MEAN, SDEViation, MAXimum, MINimum, PKPK,
NONE.
Query math format.
Enable (1 or ON) or disable (0 or OFF) calculation.
Query state of math function.
Recalculate raw input data in buffer.
Read math result of CALC 2.
:CALCulate3
:LIMit[1]
:UPPer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
:LOWer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
:STATe <b>
:STATe?
:FAIL?
:CLEAR
[:IMMediate]
:AUTO <b>
:AUTO?
:LIMit2
:UPPer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
Subsystem to control CALC 3 (limit tests):
Path to control LIMIT 1 test:
Path to configure upper limit:
Specify upper limit (-9.99e35 to +9.99e35).
Query upper limit.
Specify digital output (0 to 15).
Query source.
Path to configure lower limit:
Specify lower limit (-9.99e35 to +9.99e35).
Query lower limit.
Specify digital output (0 to 15).
Query source.
Enable (1 or ON) or disable (0 or OFF) limit test.
Query state of limit test.
Query test result (1=pass, 0=fail).
Path to clear failed test:
Clear failed test indication.
Enable (1 or ON) or disable (0 or OFF) auto clear.
Query auto clear.
Path to control LIMIT 2 test:
Path to configure upper limit:
Specify upper limit (-9.99e35 to +9.999999e35).
Query upper limit.
Specify digital output (0 to 15).
Query source (digital output value).

Table of Contents

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
BrandKeithley
Model2001
CategoryMultimeter
LanguageEnglish

Related product manuals