5-12
5.4.7 Removing the LCD Screen
CAUTION
z Do not touch the LCD screen.
z Disassemble the LCD screen in an environment as dust-free as possible.
1. Unscrew the eight M3×6 screws of T6 or T8 monitor and take the screen assembly out
carefully. To prevent the screen from being contaminated by dust, do not touch the screen.
Unscrew the twelve M3×6 screws of T9 monitor and take the screen assembly out carefully.
To prevent the screen from being contaminated by dust, do not touch the screen.