3-24
3.4.12 BIS Test
You can ch o o se eith e r of t he foll o wi ng met h ods t o perf o rm the t est:
Method 1:
Tools required:
None.
1. Connect the BIS sensor to a healthy, wide-awake adult as directed in the Operator’s
Manual.
2. Check the EEG wave and BIS numerics displayed on the screen and make sure the BIS
value is within 80-100.
Method 2:
Tools required:
BIS simulator (with flexible cable)
1. Connect the BIS sensor with the BIS simulator and select [BIS Setup] [BIS Sensor
Check] to perform a cyclic impedance check.
2. After the cyclic impedance check is finished, check that the result for each electrode is
pass.
3. Check the EEG wave and BIS numeric displayed on the screen.