A-12
A.10 Electrical Safety Inspection after Repair
The following table specifies test items to be performed after the equipment is repaired.
Repair with main unit not disassembled
Test items: 1, 2, 3
When neither power supply
PCBA nor patient
electrically-connected PCBA is
repaired or replaced
Test items: 1, 2, 3, 4
When power supply PCBA is
repaired or replaced
Test items: 1, 2, 3, 4, 5
When patient
electrically-connected PCBA is
repaired or replaced
Test items: 1, 2, 3, 4, 6, 7, 8
Repair with
main unit
disassembled
When both power supply PCBA
and patient
electrically-connected PCBA are
repaired or replaced
Test items: 1, 2, 3, 4, 5, 6, 7, 8