EasyManuals Logo
Home>Mindray>Monitor>BeneView T8

Mindray BeneView T8 Service Manual

Mindray BeneView T8
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #220 background imageLoading...
Page #220 background image
A-2
A.2 Device Enclosure and Accessories
A.2.1 Visual Inspection
Test Item Acceptance Criteria
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
No residue of fluid spillage (e.g., water, coffee, chemicals,
etc.).
The enclosure and accessories
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
A.2.2 Contextual Inspection
Test Item Acceptance Criteria
No unusual noises (e.g., a rattle inside the case).
No unusual smells (e.g., burning or smoky smells,
particularly from ventilation holes).
The enclosure and accessories
No taped notes that may suggest device deficiencies or
operator concerns.
A.3 Device Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and
legible.
Main unit label
Integrated warning labels
A.4 Protective Earth Resistance
Protective Earth Resistance is measured using the RED test lead attached to the DUT
Protective Earth terminal or enclosure. Select the test current (25 amp) by pressing SOFT
KEY 3. The front panel outlet power is turned off for this test.
The following conditions apply: L1 and L2 Open.

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals