EasyManuals Logo
Home>Mindray>Medical Equipment>DC-70 Exp

Mindray DC-70 Exp User Manual

Mindray DC-70 Exp
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #223 background imageLoading...
Page #223 background image
Care and Maintenance 12-1
12 Care and Maintenance
12.1 Overview
The maintenance procedure in this chapter is only for recommendation.
12.1.1 Tools, Measurement Devices and Consumables
Table 12-1 The list for Tools and Measurement Devices
Tools/Measurement
Devices
Qty. Remarks
Plastic and resin
container
1
Used to contain the physiological saline and two
probes available in the container.
Soft brush 1
The size is similar to that of the brush. The
material of the brush should be soft.
Small plastic bowl 1 Used to contain soap-suds
Safety testing device 1 See Appendix A for details
Table 12-2 Consumables List
12.1.2 Routine Maintenance Items
Table 12-3 The List for Maintenance Items and Maintenance Frequency
No. Item Frequency Method
1.
Dust-proof cover cleaning 1 time/month See Chapter 12.2.1
2.
The monitor and the touch screen
cleaning
1 time/month Ditto
Consumable Qty. Remarks
Aluminum foil About 1 meter long /
Physiological saline About 1000 mL
Reach the half of the container to submerge
the probe in the saline.
(See Appendix A for details)
(Concentration 0.85% to 0.95%)
Mild soap-suds About 400 mL /
Dry soft cloth 5 pieces /

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-70 Exp and is the answer not in the manual?

Mindray DC-70 Exp Specifications

General IconGeneral
BrandMindray
ModelDC-70 Exp
CategoryMedical Equipment
LanguageEnglish

Related product manuals