Care and Maintenance 12-9
12.3.3 Check for Peripherals and Optional Functions
If there is no relevant module or option in the system configuration, the relevant check can be
ignored.
Table 12-7 Peripheral and option checking list
Item Method
1.
Color/Black/White
video
printer
Check whether the output of video printer works well.
2.
Graph/text printer
Check whether the output of graph/text printer works well.
3.
Footswitch
Check whether the footswitch works as the system configures.
4.
DVD-R/W
Check whether DVD-R/W works well (burning, read/write and
openness).
5.
Barcode reader
Check whether the reader works well and whether the output is
6.
DICOM
Check whether DICOM works well, and verify whether sending
images to DICOM server via the shortcut key is normal.
7.
ECG module
Check user’s basic operation. Verify the implementation of ECG
module.
See Chapter 5.3.
12.3.4 Mechanical Safety Inspection
Mechanical safety inspection is mainly used to check mechanical strength and mechanical function
of the key assembly of ultrasonic system. The mode of test evaluation mainly is: Perform the
evaluation by means of visual check and operating check, if the check result cannot pass, the
system is in abnormal status now. Stop using the system and adopt proper measures. The test flow
is as following: