EasyManuals Logo
Home>Tektronix>Test Equipment>453

Tektronix 453 User Manual

Tektronix 453
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
Intensity
modulation
-
-
- -
-
~-
-
-
.
.
(A)
Intensity
modulation
~--
--·
·--
--~
:-~---
--
·--
.
.
(BJ
Fig.
5-50.
(Al
Typical
CRT
display
when
checking
Z-axis
operation
at
50
kilohertz,
(Bl
typical
CRT
display
when
checking
Z-axis
opera-
tion
at
50
megahertz.
d.
Set the A
SOURCE
to
EXT.
e.
Remove
the ground strap
from
between the binding
posts.
:f.
Set the constant-amplitude generator for five volts
output
at
50 kilohertz
(use
calibrated position of generator
amplitude control).
g. CHECK-CRT display for noticeable intensity modula-
tion (see
Fig.
5-50A).
The
INTENSITY
control setting may
need to be reduced to view trace modulation.
h.
Set the constant-amplitude generator for
five
volts
output
at
50 megahertz
(use
calibrated position of generator
amplitude control).
1.
Set the A and B
TIME/DIV
switch to
.1
µs.
j.
CHECK-CRT display for noticeable intensity modu-
lation (see
Fig.
5-508).
The
INTENSITY
control setting may
need to be reduced to view trace modulation.
k.
Disconnect all test equipment and replace ground strap.
©
Performance Check/
Calibration-Type
453/R453
Zero
-
...-
volt_..
- ·--
I
level
I
I
I I
I
I
I
I
I
I
I
I
I
I I
s,..__
Duration--about
_..
I
5.5
divisions
I
t--
Amplitude-2.4
divisions,
-+-0.24
-
division
I
It
I
------
Fig.
5-51.
Typical
CRT
display
when
checking
gate
amplitude
and
duration
(vertical
deflection,
five
volts/division;
sweep
rate,
two
milli-
seconds/
division
l.
73. Check A Gate Output Signal
a. Set the A and B
TIME/DIV
switch to 1
ms.
b.
Connect the A
GATE
connector
(on
side panel) to the
test-oscilloscope input connector with the 42-inch
BNC
cable.
c.
Set the test oscilloscope for a vertical deflection fac-
tor of five-volts/division
at
a sweep rate of two milliseconds/
division.
d.
CHECK-Test oscilloscope display for
2.4
divisions,
-+-0.24
division, vertical deflection
with
the bottom of the
waveform near the zero-volt level
(12
volts,
-+-10%;
see
Fig.
5-51).
Gate
duration should be about 5.5 divisions (about
11
times the A
TIME/DIV
switch setting).
7 4. Check B Gate Output Signal
a. Connect the B
GATE
connector (side panel) to the
test-oscilloscope input connector with the 42-inch
BNC
cable.
b.
Change the following control settings:
A
TIME/DIV
B
TIME/DIV
HORIZ
DISPLAY
2
ms
1
ms
DELAYED
SWEEP
(B)
c.
CHECK-Test oscilloscope display for
2.4
divisions,
-+-0.24
division, vertical deflection with the bottom of the
waveform near the zero-volt
level
(12
volts, ± 10%; see
Fig.
5-51).
Gate
duration should be about 5.5 divisions (about
11
times the B
TIME/DIV
switch setting).
This
completes the calibration procedure for the Type
453.
Disconnect all test equipment and replace the top and bot-
tom
covers.
If
the instrument has been completely checked
and calibrated to the tolerances given
in
this
procedure,
it
will
meet the electrical characteristics listed
in
the
Per-
formance Requirement column of the Specifications section
of
this
manual.
5-43

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 453 and is the answer not in the manual?

Tektronix 453 Specifications

General IconGeneral
BrandTektronix
Model453
CategoryTest Equipment
LanguageEnglish

Related product manuals