Removal and Installation Procedures
AFG3000 Series Arbitrary/Function Generators Service Manual 6-11
1. Remove the protective backing from the adhesive urethane foam gaskets
(348-1848-00 x 2, 348-1849-00 x 2, and 348-1850-00 x 1).
2. Attach the gaskets firmly to the inside of the case as shown in Figure 6-4,
making sure that the gaskets stick out 2.0 mm to 2.5 mm from the edge of the
case to ensure good contact with chassis ground.
Figure 6-4: Pasting the urethane foam gaskets
NOTE. If your instrument is an AWG310x or AFG325x, urethane foam gaskets must
be installed in the case. Follow the steps below.
348-1850-00
348-1849-00
348-1848-00
1 mm
1 mm
2.0 to 2.5 mm
348-1849-00
Gasket
Rear case