EasyManuals Logo
Home>Tektronix>Portable Generator>AFG3102

Tektronix AFG3102 Service Manual

Tektronix AFG3102
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
Performance Verification
4-24 AFG3000 Series Arbitrary/Function Generators Service Manual
3. Connect the arbitrary/function generator to the DMM as shown in Figure 4-5.
Be sure to connect the 50 Ω terminator to the arbitrary/function generator
Output connector side.
Figure 4-5: DC offset tests
4. Set up the arbitrary/function generator using the following steps:
5. Verify that each offset measurement is within the range specified in the
following tables.
Select menu Setting Operation
Function DC More... (front) > More Waveform Menu > DC
Offset + 5.000 V
+ 2.500 V (AFG325x)
Offset/Low (front)
Channel 1 Output On On (front)
AFG3011 Measurement Range
Function Offset
DC + 5.000 Vdc Vdc (5.000 × CF ± 0.110) Vdc
DC 0.000 Vdc Vdc ± 0.010 Vdc
DC – 5.000 Vdc Vdc (– 5.000 × CF ± 0.110) Vdc
DMM
50 Ω BNC coaxial cable + 50 Ω terminator
CH1 Output
AFG3000 series
BNC-Dual banana
adapter