Brief Procedures (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
17. Rotate the Vert
ical Position knob counter clockwise.
18. Verify that the displayed position of channel 1 moves lower on the display.
19. Rotate the Vertical Position knob clockwise, back to its original setting.
20. Pull down the Digital menu and select Digital Setup.
21. Double click the D0 Position and enter -1.66 and p ress Enter.
22. Verify that the d isplayed position of d igital channel moves up on the display.
23. Repeat ste
ps 1 through 22, replacing D0 with the digital channel number of
the next channel you want to verify.
24. Remove th
e test hookup: Disconnect the test hookup from the instrument.
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series 3–17