EasyManuals Logo
Home>Tektronix>Test Equipment>DPO70000 Series

Tektronix DPO70000 Series Technical Reference

Tektronix DPO70000 Series
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #113 background imageLoading...
Page #113 background image
Brief Procedures (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
17. Rotate the Vert
ical Position knob counter clockwise.
18. Verify that the displayed position of channel 1 moves lower on the display.
19. Rotate the Vertical Position knob clockwise, back to its original setting.
20. Pull down the Digital menu and select Digital Setup.
21. Double click the D0 Position and enter -1.66 and p ress Enter.
22. Verify that the d isplayed position of d igital channel moves up on the display.
23. Repeat ste
ps 1 through 22, replacing D0 with the digital channel number of
the next channel you want to verify.
24. Remove th
e test hookup: Disconnect the test hookup from the instrument.