Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
d. Display the tes
t signal:
From the tool bar touch Vertical and then touch Position.
Use the keypad to set vertical position to the setting listed in the table.
Touch Offset.
Use the keypad to set vertical offset to the positive-polarity setting
listed in the table for the current vertical scale setting. The baseline
level may move off screen.
Set the generator to the level and polarity indicated in the table for the
vertical scale, position, and offset settings you have made.
e. Measure the test signal: Press Close. Read the measurement results at the
Mean measurement readout. Refer to the following gure.
Figure 3-12: Measurement of offset accuracy
3–58 MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series