Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
Figure 3-8: Initial test hookup
2. Confirm input channels are within limits for DC accuracy at maximum offset
and position: Do the following substeps - test Ch 1 first, skipping substep aof
this step since Ch 1 is already selected from step 1.
a. Select an unchecked channel:
From the button bar, touch Measure and then Clear All to remove
the previous measurement.
Press the Vertical button of the channel just confirmed to remove the
channel from the display.
Press the front-panel Vertical button that corresponds to the channel
you are to confirm.
<4GHzmodels: TouchVe rtica l, select Vertical Setup,andthen
touch Termination 50 Ω.
Set the generator output to 0 V.
Move the test hookup to the channel you selected.
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series 3–43