Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
d. Display the tes
t signal:
From the button bar touch Vertical and touch Position.
Use the keypad to set vertical position to a position setting listed in
the t able for the current vertical scale setting. The baseline level may
move off scr
een.
Touch Offset.
Use the keypad to set vertical offset to the positive-polarity setting
listed in the table for the current vertical scale setting. The baseline
level will
remain off screen.
Set the generator to the level and polarity indicated in the table for the
vertical
scale, position, and offset settings you have made.
e. Measure the test signal: Press the X (close) button. Read the measurement
results
at the measurement statistics Mean measurement readout. See
the following gure.
Fi
gure 3-9: Measurement of DC accuracy at maximum offset and position
3–46 MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals