Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
f. Check against l
imits:
CHECK that the readout for the measurement Mean readout on
screen is with
in the limits listed for the current vertical scale and
position/offset/generator settings. E nter the value on test record.
Repeat subs
tep d, reversing the polarity of the position, offset, and
generator settings as is listed in the table.
CHECK that t
he Mean measurement readout on screen is
within the limits listed for the current vertical scale setting and
position/offset/generator settings. E nter the value on test record.
Repeat substeps c through f until all vertic al scale settings are checked
for the channel under test. (See Table 3-5.)
g. Test all channels: Repeat step 2 substeps a through f for all four channels.
3. Disconn
ect the h ookup:
a. Set the generator output to 0 V.
b. Disconnect the equipment from the generator output and the input
connector of the channel last tested.
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series 3–47