Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
Performance T
ests (MSO70000/C Series,
DSA/DPO70000B/C Series, and DPO7000 Series)
This section contains a collection of manual procedures for checking that the
instrument performs as warranted.
The procedures are arranged in logical groupings: Signal Acquisition System
Checks, Time Base System Checks, Triggering System Checks, Output Ports
Checks,an
d Serial Trigger Checks. They check all the characteristics that are
designated as checked in Specications. (The characteristics that are checked
appear with a
in Specications).
STOP. These procedures extend the condence level provided by the basic
procedures. The basic procedures should be done rst, then these procedures
performed if desired.
Prerequisites
The tests in this section comprise an extensive, valid conrmation of performance
and functionality when the following requirements are met:
The cabinet must be installed on the instrument.
Yo u must have performed and passed the procedures under Self Tests,and
those under Functional Tests. (See page 3-5, Functional Tests.)
A signal-path compensa tion must have been done within the recommended
calibration interval and at a temperature within ±5 ºC (±9 ºF) of the present
operating temperature. (If at the time you did the prerequisite Self Tests, the
temperature was within the limits just stated, consider this prerequisite met).
As
ignal-path compensation must have been done at an ambient humidity
within 25% of the current ambient humidity and after having been at that
humidity for at least 4 hours.
The instrument must have been last adjusted at an ambient temperature
between +18 ºC (+64 ºF) and +28 ºC (+82 ºF), must have been operating for a
warm-upperiodofatleast20minutes,andmustbeoperatingatanambient
temperature as listed in the specications. (See Table 1-10 on page 1-58.)
(The warm-up requirement is usually met in the course of meeting the Self
Tests and Functional Tests prerequisites listed above).
Support sensor, p robe, and adapter setups to avoid stress or torque when
connected to the device under test (DUT).
3–18 MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals