Appendix E: Performance Verification
This appendix contains performance verification procedures for the
specifications marked with the symbol. The following equipment,
or a suitable equivalent, is required to complete these procedures.
Description
Minimum
requirements
Examples
DC Voltage Source 3 mV to 4 V, ±0.1%
accuracy
a
k
500
Oscilloscope Calibrator
Leveled Sine Wave Gen-
erator
50 kHz to 500 MHz,
±3% amplitude accuracy
w
h
w
510
u
pu
Modules
Time Mark Generator 10 ms period, ±50 ppm
accuracy
50 W feedthrough
termination
BNC connectors Tektronix part number
011-0099-00
You may need additional cables and adapters, depending on the
actual test equipment you use.
These procedures cover all TDS3000 oscilloscope models. Please
disregard checks that do not apply to the specific model you are
testing.
Photocopy the test record on the next three pages and use them to
record the performance test results for your oscilloscope.