Getting Started
1–4
TDS3000 Series User Manual
Signal Path Compensation (SPC)
The signal path compensation (SPC) routine optimizes the
oscilloscope signal path for maximum measurement accuracy. You
can run the routine anytime but you should always run the routine if
the ambient temperature changes by 10 C or more.
To compensate the signal path, do the following steps:
1. Disconnect any probes or cables from the channel input
connectors.
2. Push the UTILITY button.
3. Push the System screen button to select Cal.
4. Push the Signal Path screen button.
5. Push OK Compensate Signal Path. This procedure takes several
minutes to complete.
NOTE. The signal path compensation does not include calibration to
the probe tip.
Product and Feature Description
The TDS3000 Series Digital Phosphor Oscilloscope family consists
of the models in the table below.
Model Bandwidth
Maximum
sample rate
TDS3012 (2 Ch), TDS3014 (4 Ch) 100 MHz 1.25 GS/s
TDS3032 (2 Ch), TDS3034 (4 Ch) 300 MHz 2.5 GS/s
TDS3052 (2 Ch), TDS3054 (4 Ch) 500 MHz 5 GS/s

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals