Appendix E: Performance Verification
TDS3000 Series User Manual
E–3
Performance checks High limitTest resultLow limit
hann
2
1 mV/div 99.25 mV 100.8 mV
D
measurement
2 mV/div –7.540 mV –6.460 mV
accuracy
5 mV/div
–101.8 mV –98.24 mV
50 mV/div 982.4 mV 1.018 V
50 mV/div 632.4 mV 667.6 mV
50 mV delta 340.5 mV 359.5 mV
90 mV/div –339.3 mV –290.7 mV
200 mV/div 9.900 V 10.10 V
1 V/div –10.30 V –9.698 V
hann
3
1 mV/div 99.25 mV 100.8 mV
D
measurement
2 mV/div –7.540 mV –6.460 mV
accuracy
5 mV/div
–101.8 mV –98.24 mV
50 mV/div 982.4 mV 1.018 V
50 mV/div 632.4 mV 667.6 mV
50 mV delta 340.5 mV 359.5 mV
90 mV/div –339.3 mV –290.7 mV
200 mV/div 9.900 V 10.10 V
1 V/div –10.30 V –9.698 V