Message Possible cause Suggested action Alarm priority
Tube type does not
match device
configuration.
The tube type settings and the
actual tube type do not match
Change the tube type setting to
match the actual tube type.
Information
SpO2 messages
General SpO2 messages
Message Possible cause Suggested action Alarm priority
SpO2 not functional.
044800
The SpO2 power on self test
failed
Internal hardware malfunction in
SpO2 module. Replace the module.
Very low
SpO2 rebooting. 044900 The SpO2 module is rebooting Informational error. The host
software is attempting to clear an
error by rebooting the SpO2 module.
No action required.
Very low
SpO2 rebooting. 044901 The SpO2 module stopped
sending data
Informational error. The host
software is attempting to clear an
error by rebooting the SpO2 module.
No action required.
Very low
SpO2 rebooting. 044902 SpO2 received a packet with bad
CRC from module
Informational error. The host has
received a packet with bad CRC from
the SpO2 module. The packet in
question is ignored. No action
required.
Very low
SpO2 rebooting. 044903 The SpO2 power on self test
failed
Internal hardware malfunction in
SpO2 module. Replace the module.
Very low
SpO2 rebooting. 044904 The SpO2 power on self test
timed out
Internal hardware malfunction in
SpO2 module. Replace the module.
Very low
Masimo messages
Message
Possible cause Suggested action Alarm priority
Sensor not connected.
Clear to retry. 040600
The SpO2 cable is not connected Connect SpO2 cable. If the problem
is still present, replace the SpO2
sensor. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
Replace the SpO2 cable.
040700
The SpO2 cable life has expired Replace the SpO2 cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
Very low
Directions for use Troubleshooting 95

Table of Contents

Other manuals for Welch Allyn Connex Spot Monitor

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandWelch Allyn
ModelConnex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals