Software update messages
Message Possible cause Suggested action
Alarm
priority
Software Update:
Manifest transfer timed
out. Verify connection
and retry.
The manifest file transfer timed
out or connection was lost during
download
Verify the connection and retry. Information
Software Update:
Package file transfer
timed out. Verify
connection and retry
The package file transfer timed
out or connection was lost during
download
Verify the connection and retry. Information
Software Update:
Invalid token file.
There was an invalid token file Verify and update the token file. Information
Software Update:
Unable to find manifest
file on server.
The manifest file was not found
on the server
Verify the manifest file is on the server. Information
Software Update:
Unable to verify
manifest file signature.
The manifest file signature
verification failed.
Regenerate the software package and retry. Information
Software Update:
Package file corrupted.
Regenerate package and
retry.
The package file is corrupt, does
not have expected SHA256 hash
Regenerate the software package and retry. Information
Software Update:
Unable to find package
file.
The package file cannot be found Verify the package file is on the server. Information
Software Update:
Installation failed.
Reboot and retry.
At least one of the sub systems
failed to install
Restart the monitor. Information
Software Update:
Upgrade unsuccessful.
Insufficient disk space.
The partition is running out of
space
Free up adequate space needed to perform
the upgrade.
Information
Software Update:
Update unsuccessful.
Incompatible firmware.
The current firmware version is
lower than the one required by the
token file
Try to update to an earlier software
package.
Information
Software Update: SWUP
internal error
SWUP NIBP is not functional Information status message; click OK button
to dismiss.
Information
Software Update:
Manager internal error
The Software Update manager is
not functional
Information status message; click OK button
to dismiss.
Information
Radio software upgrade
failed. 350600
The radio software was not
upgraded.
Check for software update and apply it. If
message is still present, replace the radio.
Very low
126 Troubleshooting Connex
®
Spot Monitor

Table of Contents

Other manuals for Welch Allyn Connex Spot Monitor

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandWelch Allyn
ModelConnex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals