Message Possible cause Suggested action Alarm priority
SpO2 interference
detected. Clear to retry.
040400
The SpO2 interference detected. Reapply the sensor to the patient. If
the problem is still present, replace
the SpO2 Sensor. If the problem is
still present, replace the cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists,
replace the module.
Very low
Low SpO2 perfusion
index. Clear to retry.
040500
SpO2 marginal pulse quality or
artifact
Reapply the sensor to the patient. If
the problem is still present, replace
the SpO2 Sensor. If the problem is
still present, replace the cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists,
replace the module.
Very low
1
Pulse search is a normal part of operation and does not have an associated corrective action.
Temperature messages
SureTemp messages
Message
Possible cause Suggested action Alarm priority
Temperature not functional.
30105
WACP message CRC
mismatch on
temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30201
This message is not
implemented by the
temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30202
This message is not
supported by the
temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30203
The temperature
module has run out of
memory.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30205
The temperature
module received a
invalid parameter
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30206
The parameter provided
by the temperature
module is outside of
the allowable range for
the specified message.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30207
The temperature
module message
Internal malfunction. If the problem
persists, replace the module.
Very low
106 Troubleshooting Connex
®