Note Modifying the date and time on the device modifies the date
and time on all saved readings on the device.
Modifiers
The Modifiers tab enables you to enter additional information for current measurements.
Set modifiers
1. On the Home tab, press and hold the parameter desired.
The Modifiers screen appears.
2. Touch the desired parameter on the Modifiers screen and use the keypad for manual
entry of NIBP, SpO2, Pulse Rate, Temperature, or Additional parameters.
3. Touch OK to accept the entry.
4. Touch OK to accept the changes and return to the Home tab or touch Cancel to
delete all entries.
The Modifier settings clear after a power cycle, after you clear or save the Home tab,
or after you select a new patient.
Patient list
From the Patient List screen, you can do the following:
•
Retrieve a patient list from the network (for example, electronic medical records
(EMR) or a central station).
• Select a patient from the list.
• Manually create a patient record.
• Search Patient List
Warning Patient injury risk. Verify patient identity on the monitor after
manual or barcode entry and before printing or transferring patient records.
Failure to identify the correct patient can result in patient injury.
Select a patient
The options for selecting previously stored patients from the List tab vary based on the
following conditions:
•
Active profile
• Established patient context
• Connection to a network
• Connection to a central station
Based on the boldface text presented, follow the steps below that apply to your patient
and the device.
1.
In all profiles but Office, when patient context has not been established on the
device:
a. Touch the Patient tab.
Directions for use Patient data management 45