7-83
IM AQ6370C-17EN
Remote Commands
1
2
3
4
5
6
7
8
App
Index
SENSe Sub System Command
:SENSe:AVERage:COUNt
Function Sets/queries the number of times averaging for
each measured point.
Syntax
:SENSe:AVERage:COUNt<wsp><integer>
:SENSe:AVERage:COUNt?
<integer> = Number of times averaging
Example:
:SENSE:AVERAGE:COUNT 100
:SENSE:AVERAGE:COUNT? -> 100
Explanation This is a sequential command.
:SENSe:BANDwidth|:BWIDth[:RESoluti
on]
Function Sets/queries the measurment resolution.
Syntax
:SENSe:BANDwidth|:BWIDth[:RESolutio
n]<wsp><NRf>[M|Hz]
:SENSe:BANDwidth|:BWIDth
[:RESolution]?
<NRf> = Measurement resolution [m|Hz]
Response
<NRf> [m|Hz] (AQ6370C/AQ6370D/AQ6373B/
AQ6373)
<NRf> [m|Hz|m
-1
] (AQ6375/AQ6375B)
Example
:SENSE:BANDWIDTH:RESOLUTION 20PM
:SENSE:BANDWIDTH?
-> +2.00000000E-012
Explanation For the AQ6375/AQ6375B, to set using the
wavenumber, do not add any units when
in Wavenumber mode, and just enter the
numerical value.
This is a sequential command.
:SENSe:CHOPper
Function Sets/queries chopper mode.
Syntax
:SENSe:CHOPper<wsp>OFF|SWITch|0|2
:SENSe:CHOPper?
Response 0 = OFF,
2 = SWITCH
Example
:SENSE:CHOPPER SWITCH
:SENSE:CHOPPER? -> 2
Explanation When the measurement sensitivity setting
(
:SENSe:SENSe
command) is NORMAL
HOLD or NORMAL AUTO, Chopper does not
function even if chopper mode is turned on
with this command.
With the AQ6375/AQ6375B, the command is
invalid.
This is a sequential command.
:SENSe:CORRection:LEVel:SHIFt
Function Sets/queries the offset value for the level.
Syntax
:SENSe:CORRection:LEVel:SHIFt<wsp><
NRf>[DB]
:SENSe:CORRection:LEVel:SHIFt?
<NRf> = Level offset value [dB]
Example
:SENSE:CORRECTION:LEVEL:SHIFT 0.2DB
:SENSE:CORRECTION:LEVEL:SHIFT?->
+2.00000000E-001
Explanation This is a sequential command.
:SENSe:CORRection:RVELocity:MEDium
Function Sets/queries whether air or vacuum is used as
the wavelength reference.
Syntax
:SENSe:CORRection:RVELocity:MEDium
<wsp>AIR|VACuum|0|1
:SENSe:CORRection:RVELocity:MEDium?
AIR = Air is assumed to be the reference.
VACuum = Vacuum is assumed to be the
reference.
Response 0 = AIR
1 = VACuum
Example
:SENSE:CORRECTION:RVELOCITY:
MEDIUM VACUUM
:SENSE:CORRECTION:RVELOCITY:MEDI
UM?-> 1
Explanation This is a sequential command.
:SENSe:CORRection:WAVelength:SHIFt
Function Sets/queries the offset value for the
levelwavelength.
Syntax
:SENSe:CORRection:WAVelength:SHIFt<
wsp><NRf>[M]
:SENSe:CORRection:WAVelength:SHIFt?
<NRf>= Wavelength offset value [m]
Example
:SENSE:CORRECTION:WANELENGTH:
SHIFT 0.05NM
:SENSE:CORRECTION:WANELENGTH:SHI
FT?-> +5.00000000E-011
Explanation This is a sequential command.
:SENSe:SENSe
Function Sets/queries the measurement sensitivity.
Syntax
:SENSe:SENSe<wsp><sense>
:SENSe:SENSe?
<sense>= Sensitivity setting parameters
NHLD = NORMAL HOLD
NAUT = NORMAL AUTO
NORMal = NORMAL
MID = MID
HIGH1 = HIGH1 or HIGH1/CHOP
HIGH2 = HIGH2 or HIGH2/CHOP
HIGH3 = HIGH3 or HIGH3/CHOP
Response 0 = NHLD
1 = NAUT
2 = MID
3 = HIGH1
4 = HIGH2
5 = HIGH3
6 = NORMAL
Example
:SENSE:SENSE MID
:SENSE:SENSE? -> 2
Explanation This is a sequential command.
7.6 Instrument-Specific Commands

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AQ6370C and is the answer not in the manual?

YOKOGAWA AQ6370C Specifications

General IconGeneral
BrandYOKOGAWA
ModelAQ6370C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals