D98 User Guide
Resetting source odometers
While the CoreSense M10 is in operation,
persistent files
are created for each source, describing certain
parameters and source states. These files are designed to remain in the instrument. However, there are
times when you need to erase those files (e.g., after replacing a source).
To do so:
1 Click Source diagnostics from the Maintenance page. The Source diagnostics page appears (see
Figure 68 on page D97).
2 Click Reset odometers.
The source persistent files are deleted, allowing the system to use all sources in a fresh state again.
NOTICE
Clicking the Reset odometers button by mistake will delete the persistent files and
recreate them based on the current instrument state.
Reinitializing source initial data
Upon installation, spectrum files are created based on the sources used. Should you replace a source
for any reason, you will have to initialize these spectrum files.
To do so:
1 Click Source diagnostics from the Maintenance page. The Source diagnostics page appears (see
Figure 68 on page D97).
2 Click Reinitialize initial data. The spectrum files are initialized and recreated immediately by the
CoreSense M10 system.
Managing Coresense M10 configurations
Activation and backup of CoreSense M10 configurations is managed from the Configuration page on
the Maintenance page.
—
Figure 69 The Configuration page