417102012_LMI02.doc - 68 - Rev. 9-07.08
7 Settings and Startup
7.1 Operating level
At the operating level, you can:
• Start and stop pre-metering with the arrow keys (in AS mode)
• Display the uncompensated conductivity value for control purposes.
• Switch off the instrument.
From the operating level, you can enter the code to switch the instrument into the
service, parameter and configuration levels.
7.1.1 Key functions
fig 7.1
Operating level Key(s) Service, parameter and
configuration levels
Activation of menu:
Access code
NEXT
Call up a menu. Return to starting
level
Display uncompensated conductivity
value with pressed key (with a *
before the conductivity value)
Selection of digit of access code
↵ (SEL)
Call up menu, scroll in the menu,
move the cursor
Increase value of access code
Start pre-metering (press key 5s)
Increase the value in a menu
Decrease value of access code
Stop pre-metering (press key 5s)
Switch to the next display in the
information level.
Decrease the value in a menu
Switch off instrument
Press any key to restart the
instrument.
and and ↵ (press
simultaneously for 5
seconds
Switch to information level.
(Display of limit values SP1 and
SP2, output current and Tkα as well
as pre-metering and max. metering
time)
and
(press together briefly)