7.3 Parameter level
7.3.1 Settings
The following settings can be performed in the parameter level
(PARAMETER FUNCTIONS):
• ADJUSTMENT FUNCTIONS
- Adjust the conductivity display (CONDUCTIVITY +/- 0.0 % = XXX mS/cm)
- Adjust the temperature display (TEMPERATURE XX °C)
• DEFINE METERING SETTINGS (factory setting in brackets)
- Set pre-metering time (PREMETERING TIME: 0120 s)
- Set pre-metering delay (PREMETERING DELAY: 000 s)
- Set concentration of SP1 (CONZENTRATION SP1: OFF mS/cm)
- Set concentration of SP1 (CONZENTRATION SP2: OFF mS/cm)
- Set temperature compensation for alkali or acid detergent
(TEMPERATURE COMP. ALKAL ACID)
• ALARM CONTROL SETUP
- Set delay time for detergent
empty message (EMPTY DETERGENT DELAY: 0005 s)
- Set maximum metering time (DETERGENT DELAY TIME: 0180 s)
- Set manual metering time (MANUAL DOSING TIME: 0120 s)
- Set concentration overdose of detergent (DET. OVERDOSAGE CONZ.: 100 mS/cm)
- Set delay time for detergent concentration overdose message
(DET: OVERDOSE DELAY: 120 s)
- Set lower limit for temperature (TEMPERATURE MINIMUM: OFF)
- Set delay time for temperature lower limit alarm (only shown if temperature value set)
(TEMPERATURE DELAY: 0120 s)
- Set repeat alarm (ALARM REPEAT TIME: OFF)
• SET PARAMETERS TO DEFAULT
- Resets parameters to the factory settings (ARE YOU SURE?)
NOTE
The programming steps and display texts described in the following sections of this
chapter are summarised in the “Programming Structure” overview in the Appendix.