4.6.5 Connect the nitrogen supply
WARNING: DANGEROUS SUBSTANCES
Risk of asphyxiaon. A r
elease of nitrogen can cause asphyxiaon. The nitrogen
supply must enable isolaon in accordance with SEMI S2 Lockout and tagout
requirements.
CAUTION: NITROGEN SUPPLY
Risk of damage to the equipment. Make sure that the nitrogen supply agrees with
requirements. If it does not, the gas pipelines can become blocked or the dry
pumping system can be damaged.
Refer to Figure: Controls/connectors on the rear of the system for locaon of the
nitrogen purge port. The gas module in the system can be adjusted to suit process
demands. Refer to Gas module conguraon on page 64.
Refer to Table: Nitrogen purge data for nitrogen supply requirements.
Flammable/pyrophoric materials
WARNING: FLAMMABLE GASES
Risk of injury and damag
e to the equipment. Obey the instrucons and make note of
precauons to make sure that the pumped gases do not enter their ammable
ranges.
When ammable or pyrophoric materials are in the pump there can be addional risks.
The user must assess and manage these risks as part of the process tool installaon.
The severity of the risks and the necessary control measures will depend on if:
▪ the tool exhaust is in the ammable region
▪ this is part of normal pr
ocess tool operaon
▪ it can only occur in rare condions.
The addional risks occur because all dry pumps must be considered a possible source
of ignion caused by the heat of compression, or fricon. If ignion occurs:
▪ High pressures can occur in the pump and possibly not be contained.
▪ A ame front can travel back up the foreline.
▪ A ame front can travel downstream from the exhaust of the pump.
Industry best pracce suggests that the following measures will decrease the risks of
pumping ammable mixtures and pyrophoric materials. The user must do a risk
assessment and take appropriate measures:
▪ Do not let air enter the equipment.
▪ Make sure that the system is leak ght.
▪ Pr
event the entry of the pump gases in the ammable range. To achieve this,
supply sucient inert gas purge to dilute the pump gases. For example, diluon
with nitrogen to below one quarter LEL (Lower Explosive Limit) or, if that is not
praccal, to below 60% LOC (Liming Oxidant Concentraon).
10/2020 - ©Edwards Limited
Page 35M56635880_D