6.4 System operang t
emperature conguraon
The system operang temperature conguraon can be adjusted to suit process
demands. Contact your supplier for instrucons on how to access the conguraon
menu (through the PDT) and for process specic recommendaons.
Use the following procedure:
1. Access the Cong menu on the PDT (requires an access code, contact supplier).
2. Scroll to Temp Set-points - press Enter.
3. Scroll to Set DP Temp - press Enter.
4. Scroll to desired temperature - Press Enter.
Note:
The Dry Pump se
t-point is the desired surface temperature of the internal low
vacuum stage.
Refer to Table: Dry pump and booster temperature sengs for default temperature
sengs and ranges.
Table 32
Dry pump and booster temperature sengs
Temperature Seng
Range in which temperature can be
set(°C)
De
fault value (°C)
Dry pump
(Standard variants)
iXM600, iXM1200
60 - 110 80
Dry pump
(Standard variants)
iXM200, iXM3000
60 - 110 110
Dry pump
(T and TX variants)
60 - 130
130
Booster
(Only TX variants)
- 85
10/2020 - ©Edwards Limited
Page 65M56635880_D
M56635880_D - Operaon