The solenoid valve(s) in the gas module close, to switch o the supply of nitrog
en
to the pump.
The solenoid valve(s) in the temperature control manifold(s) de-energise with loss
of temperature control.
The pump display terminal will display '1.01 ALARM/STOP ACTIVATED' (if
connected) or it can display ‘ALARM 186.01/ DP INV 0040 000 / EMS’.
For pXH booster pump, the PDT will display '801.01 ALARM/STOP ACTIVATED' (if
connected) or it may display ‘ALARM 825.01/ DP INV 0040 000 / EMS’.
The Running LED will go o.
The Alarm LED illuminates.
For TX variants, the exhaust heaters remain powered on.
Booster heaters on TX variants (where installed) switch o.
6.3 Gas module conguraon
6.3.1 iXM Varimode gas module adjustment procedure
The factory default gas ballast ow seng for the Varimode gas module is 44 slm.
The default low warning set-point for the Varimode gas module while gas ballast is on is
10 slm, unless the pump is in green mode.
The gas ballast seng can be adjusted by the rotaon of the gas ballast adjuster screw
accessible from the rear of the pump enclosure.
If a dierent low warning set-point other than 10 slm is required aer the variable
ballast ow seng has been changed from its default nominal 44 slm seng, then:
1. With the PDT command, select the N
2
ow w
arning menu, provided that the pump
is running and there is ow above 10 slm. Wait up to 30 seconds to make sure N
2
ow is in a steady state.
2. Aer a 5 second c
ountdown delay, the pump will use the last spot reading of the
N
2
ow rate to dene a new low and high warning threshold near to the actual
nominal ow selected. This will be stored on the parameter in the low and high
warning thresholds and will be displayed on the PDT. The value stored will be 30%
(congurable) less or more than the current ow value.
Use PDT up and down keys to change the low warning threshold in a range of
30% less than curren
t nominal N
2
ow (with a minimum low warning
threshold of 5 slm).
If the seng is aempt
ed when a varimode gas module is not present, there is
no valid ow reading at the start, then the PDT will display: 'CANNOT SET NO
READING'.
If the ow uctuates by more than 10 slm during the 5 second countdown
delay, then the PDT will display: 'CANNOT SET FLOW CHANGING'. The
countdown will then be cancelled and the seng is aborted.
If the ow reading goes beneath a value where calculang 30% less than the current
ow will be below 5 slm, then the PDT will display 'CANNOT SET LOW READING'. The
countdown will then be cancelled and the seng aborted.
10/2020 - ©Edwards Limited
Page 64M56635880_D
M56635880_D - Oper
aon

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards iXM Series and is the answer not in the manual?

Edwards iXM Series Specifications

General IconGeneral
BrandEdwards
ModeliXM Series
CategoryWater Pump
LanguageEnglish

Related product manuals