Revision E MAC 5500 resting ECG analysis system 5-17
2020299-020
Maintenance: Disassembly Guidelines
the label of the system.
6. Select/verify that No is
selected for Update Primary Boot.
7. Enter the Print head resistance. This num
ber can be found on the print head
label.
8. Select the appropriate language in the Keyboard m
enu.
9. Select Retu
rn.
Restore System Setups
1. Power up the cart.
2. From the Main Menu, se
lect System Setup.
3. Enter the System Passw
ord and press Enter.
4. Select Restore S
etup from the System Setup menu.
5. Select Fr
om SD Card from the Restore Setup menu.
Restore Options
Using the option activation codes for the system, restore the options which had been
installed on the board which was removed. These options are printed on a label
located on the bottom of the paper tray.
NOTE
Use the activator codes shown on the label on your system. The activator codes
show
n in the fi
gure below are examples only.
1. Power on the cart.
2. Within the System Setup f
unction, select Basic System.
3. Select Opti
on Activation to activate options.
4. Type the 12-digit option activation code and press the Enter key.
5. Repeat the previous step for each option to be activated on the
new system.