Revision E MAC 5500 resting ECG analysis system 5-27
2020299-020
Maintenance: Functional Checkout Procedures
Visual Inspection
Inspect the following for excess wear and or any visual signs of damage.
1. Passed check for defective or broken patient cable/leadwires and out of date
electrodes?
2. Discussed electrode placement, skin prep, and patient related requirements with
ECG Tech?
3.
Keyboard/LCD passed inspection?
4. AC Power cord passed inspection?
5. Battery (swollen or leaky) passed inspection?
6. Trolley and casters passed inspection?
7. External surfaces passed inspection?
8. FUSE type and rating passed inspection?
9. Fasteners replaced and secured?
10. All harnesses’ and internal wiring has been secured?
Checkout Procedures
Perform the applicable checkout procedure.
Operational Checks
1. Power-up self-test passed?
2. Simulated recorded rhythm strip successful?
3. Simulated recorded ECG successful?
4. Simulated ECG stored on media successful?
5. Simulated ECG data transmitted successfully to MUSE?
6. Exercise device successfully communicates with MAC5500/5000?
7. Blood pressure monitor successfully communicates with
M
AC5500/5000?
8. Barcode reader successfully communicates with MAC5500/5000?
9. Magnetic card reader successfully communicates with MAC5500/5000?
10. Trolley casters lock successfully?