Revision E MAC 5500 resting ECG analysis system 3-17
2020299-020
Installation: MAC 5500 ST Requirements and Configuration
System Setup,
Exercise Test,
Inputs/Outputs, and
set Slow Analog Output to Workload.
Configure pin 8 on the ANA/TTL po
rt by selecting the following:
System Setup,
Exercise Test,
Inputs/Outputs, and
Set Fast Analog Output to Workload.
Bicycle Ergometers
Ergoline 800/900, Lode Ergometer
Connection Requirements – Use cable PN 2008109-001 (Ergoline 800), PN
2008114-001 (Ergoline 900), or PN 2007981-001 (Lode Ergometer), to connect
from the MAC 5500 ANA/TTL port to the ergometer analog control port.
NOTE
For any other ergometer, the customer is responsible for making the appropriate
cable.
Device Configuration Requirements – Refer to ergometer Operator's
Manua
l.
MAC 5500 Configuration Requirements – Use the Edit Protocol application to set
the protocol Test Type to Ergometer in Watts or Ergometer in KPM for protocols
that will be used with this ergometer.
Configure pin 2 on the ANA
/TTL po
rt by selecting the following:
System Setup,
Exercise Test,
Inputs/Outputs, and
Slow Analog Output to Workload, or
Configure pin 8 by selecting:
System Setup Exercis
e Test Inputs/Outputs
Fast Analog Output to Workload