EasyManuals Logo
Home>GE>Medical Equipment>MAC 5500

GE MAC 5500 Service Manual

GE MAC 5500
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Revision E MAC 5500 resting ECG analysis system 4-23
2020299-020
Troubleshooting: Frequently Asked Questions
Entering Patient Data
Q: Do I have to enter all the Information I see on the Patient Data screen?
A: In System Setup > Basic System >
Patient Questions you can require that the
patient identification number, or medical record number) be entered. It is not a
requirement to enter any other data. However, we recommend that you enter the
patient name and identification number, at the least. If you are transmitting to
the MUSE system you will want to enter the Location number as well. If an
emergency situation dictates that you must complete the test. without entering
the patient data, make sure you edit the record to add the missing information
before you transmit it to the MUSE system.
Transmission
Losing Fields When Transmitting
Q: Why do I lose the Referring MD and Technician names off of my reports when
I transmit records to the MUSE system?
A: Your MAC 5500 may be transmitting to the SDLC modem on the MUSE
system instead of the CSI modem. Check
in System Setup to make sure you are
transmitting to the MUSE system CSI phone number.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 5500 and is the answer not in the manual?

GE MAC 5500 Specifications

General IconGeneral
BrandGE
ModelMAC 5500
CategoryMedical Equipment
LanguageEnglish

Related product manuals