B-2 MAC™ 800 2031504-159D
Electromagnetic Compatibility
Electromagnetic Compatibility (EMC) Overview
Changes or modification to this system not expressly approved by GE
Healthcare could cause EMC issues with this or other equipment. This
system is designed and tested to comply with applicable regulation(s)
regarding EMC and needs to be installed and put into service according
to the EMC information stated as follows.
WARNING
INTERFERENCE — Use of portable phones or other
radio frequency (RF) emitting equipment near the system
may cause unexpected or adverse operation.
WARNING
PLACEMENT — The equipment or system should not be
used adjacent to, or stacked with, other equipment.
If adjacent or stacked use is necessary, the equipment or
system should be tested to verify normal operation in the
configuration in which it is being used.
Guidance and Manufacturer’s Declaration -
Electromagnetic Emissions
The MAC 800 is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to ensure
that the MAC 800 system is used in such an environment.
Emissions Test Compliance
Electromagnetic Environment -
Guidance
RF Emissions
CISPR11
Group 1 The equipment uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF Emissions
CISPR11
Class B Class B Equipment is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class B
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Complies