Maintenance
5. Press the Power key several seconds to shut down the system.
6. Press the Power key again to reboot the system.
The system has been updated.
Conditioning the Battery Pack
To maintain the storage capacity of the battery pack installed in your
MAC 800, GE Healthcare recommends that you condition your MAC 800
battery pack once every six months to reset the electronic fuel gauge
inside the battery. A condition cycle consists of an uninterrupted
“charge-discharge-charge” cycle.
You can condition the MAC 800 battery pack while installed in the MAC
800 system that is not being used to record tests on a patient.
1. Disconnect the AC mains power from the MAC 800 system.
2. Display the Battery Status window:
a. From the Main Menu, press F4 to select System Configuration.
b. Press F6 (More) > F6 (More) > F5 (Service Setup).
c. Type the service password and press F6 (OK).
d. Select the System Diagnostics button.
e. Select the Battery Test button.
The Battery Status window opens.

Table of Contents

Other manuals for GE MAC 800

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 800 and is the answer not in the manual?

GE MAC 800 Specifications

General IconGeneral
BrandGE
ModelMAC 800
CategoryMedical Equipment
LanguageEnglish

Related product manuals