Electromagnetic Compatibility
2031504-159D MAC™ 800 B-5
according to the maximum output power of the communications
equipment.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (w)
according to the transmitter manufacturer.
NOTE
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
EMC-Compliant Cables and Accessories
WARNING
USE OF CORRECT ACCESSORIES The use of
accessories, transducers and cables other than those
specified may result in increased emissions or decreased
immunity performance of the equipment or system.
The following table lists cables, transducers, and other applicable
accessories which GE Healthcare claims are EMC compliant.
NOTE
Any supplied accessories that do not affect EMC compliance are not
included.
Separation Distance in Meters (m) According to Frequency of Transmitter
Rated Maximum Output Power
of Transmitter in Watts
150 kHz to 80 MHz
outside ISM bands
d = 1.2
150 kHz to 80 MHz
in ISM bands
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01 0.12 0.12 0.12 0.23
0.1 0.38 0.38 0.38 0.73
1 1.2 1.2 1.2 2.3
10 3.8 3.8 3.8 7.3
100 12 12 12 23
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Table of Contents

Other manuals for GE MAC 800

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 800 and is the answer not in the manual?

GE MAC 800 Specifications

General IconGeneral
BrandGE
ModelMAC 800
CategoryMedical Equipment
LanguageEnglish

Related product manuals