Volara™ System, Model PVL1 User Manual (196286 REV C) 97
Specifications
POWER REQUIREMENTS
Replaceable Battery (Optional)
ENVIRONMENTAL CONDITIONS FOR USE
ENVIRONMENTAL CONDITIONS FOR TRANSPORT AND STORAGE
Condition Range
Line voltage 100 - 240 V AC
Supply frequency 50 - 60 Hz
Supply current 2.0 - 1.0 A
Mode of operation Non-continuous: 5 minutes ON/
20 minutes OFF at CHFO pressure setting
of 65 cmH2O and above.
Fuse rating Fuse, 4 A H 250 V
Feature Description
Lithium-ion Battery A new and fully charged battery can
support 6 sessions of typical Volara™
therapy or 9 sessions of typical Synclara™
therapy.
Output: 24 V DC +/- 1.5 V DC
Capacity 2700 mAh
Condition Range
Temperature 41°F to 95°F (5°C to 35°C)
ambient temperature
Relative humidity 10% to 90% non-condensing
Pressure 70 kPa to 106 kPa
Condition Range
Temperature -4°F to 140°F (-20° to 60°C)
Relative humidity 10% to 90%
Pressure 50 kPa to 106 kPa
196654__C_2019-Apr-29_Cleaned