106 Volara™ System, Model PVL1 User Manual (196286 REV C)
Specifications
ELECTROMAGNETIC IMMUNITY TO WIRELESS COMMUNICATIONS
The Volara™ System, Model PVL1 is intended for use in the electromagnetic environ-
ment specified below. The customer or user of the Volara™ System, Model PVL1 should
make sure it is used in such an environment.
Sides Tested Frequency
(MHz)
Test Severity Level Test Distance
(m)
Front, back, left, right 385 27 V/m, 50% PM 18 Hz 0.3
Front, back, left, right 450 28 V/m, FM ± 5 kHz,1 kHz 0.3
Front, back, left, right 710 9 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 745 9 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 780 9 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 810 28 V/m, 50% PM 18 Hz 0.3
Front, back, left, right 870 28 V/m, 50% PM 18 Hz 0.3
Front, back, left, right 930 28 V/m, 50% PM 18 Hz 0.3
Front, back, left, right 1720 28 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 1845 28 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 1970 28 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 2450 28 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 5240 9 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 5500 9 V/m, 50% PM 217 Hz 0.3
Front, back, left, right 5785 9 V/m, 50% PM 217 Hz 0.3